Z-0777-03

Boston Scientific Scimed Radius 3.0 mm x 31 mm Self-Expanding Coronary Stent with Over-the-Wire Delivery...

Boston Scientific Corporation / initiated 2003-03-17 / Terminated / root cause: sterility

Boston Scientific Scimed began a recall of Boston Scientific Scimed Radius 3.0 mm x 31 mm Self-Expanding Coronary Stent with Over-the-Wire Delivery System, Catalog Number H74917500301 on . The reason FDA records is "A component within the manifold of the stent delivery systems was found to have pyrogen levels above specification". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-0777-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
Boston Scientific Scimed Radius 3.0 mm x 31 mm Self-Expanding Coronary Stent with Over-the-Wire Delivery System, Catalog Number H74917500301
Product code
MAF Stent, Coronary
Reason for recall
A component within the manifold of the stent delivery systems was found to have pyrogen levels above specification.
Root cause tag
sterility
FDA root cause
Other
Action
Consignees were sent a recall letter dated March 17, 2003. The letter requested that consignees discontinue use of stents from the affected lots and return them.
Quantity in commerce
4 Stent Systems
Distribution
Nationwide in the United States
Product codes and lots
Lot number 5357125