Z-0778-03
Boston Scientific Scimed Radius 4.0 mm x 31 mm Self-Expanding Coronary Stent with Over-the-Wire Delivery...
Boston Scientific Corporation / initiated 2003-03-17 / Terminated / root cause: sterility
Boston Scientific Scimed began a recall of Boston Scientific Scimed Radius 4.0 mm x 31 mm Self-Expanding Coronary Stent with Over-the-Wire Delivery System, Catalog Number H74917500401 on . The reason FDA records is "A component within the manifold of the stent delivery systems was found to have pyrogen levels above specification". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-0778-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- Boston Scientific Scimed Radius 4.0 mm x 31 mm Self-Expanding Coronary Stent with Over-the-Wire Delivery System, Catalog Number H74917500401
- Product code
- MAF Stent, Coronary
- Reason for recall
- A component within the manifold of the stent delivery systems was found to have pyrogen levels above specification.
- Root cause tag
- sterility
- FDA root cause
- Other
- Action
- Consignees were sent a recall letter dated March 17, 2003. The letter requested that consignees discontinue use of stents from the affected lots and return them.
- Quantity in commerce
- 7 Stent Systems
- Distribution
- Nationwide in the United States
- Product codes and lots
- Lot number 5350197