Z-0870-03

Elecsys Troponin T; catalog number 2017644

Roche Diagnostics GmbH / initiated 2003-04-25 / Terminated

Roche Diagnostics Corp. began a recall of Elecsys Troponin T; catalog number 2017644 on . The reason FDA records is "Hemoglobin interference limitation revised downward; hemolyzed samples may result in false negative results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0870-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Elecsys Troponin T; catalog number 2017644
Product code
MMI Immunoassay Method, Troponin Subunit
Reason for recall
Hemoglobin interference limitation revised downward; hemolyzed samples may result in false negative results.
FDA root cause
Other
Action
An Urgent Product Correction letter was mailed to each account, during the week of 4/21/03, advising them not to use hemolyzed specimens with a hemoglobin level equal to, or greater than, 0.1 g/dL.
Distribution
United States
Product codes and lots
All lots.