Easy Moving Motor Driven Mobile X-Ray Unit, Models SM-20HF and SM-30HF; Manufactured by SEDECAL, C/Pelya 9...
Sedecal USA, Inc. began a recall of Easy Moving Motor Driven Mobile X-Ray Unit, Models SM-20HF and SM-30HF; Manufactured by SEDECAL, C/Pelya 9 Pol. Ind. Rio De Janeiro, 28110 Algete (Madrid) SPAIN on . The reason FDA records is "Incorrect identification labels and failure of the cretification testing resulting in noncompliance with the U.S> Federal performance standard". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-650/51-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sedecal USA, Inc
- Product
- Easy Moving Motor Driven Mobile X-Ray Unit, Models SM-20HF and SM-30HF; Manufactured by SEDECAL, C/Pelya 9 Pol. Ind. Rio De Janeiro, 28110 Algete (Madrid) SPAIN
- Product code
- IZL--
- Reason for recall
- Incorrect identification labels and failure of the cretification testing resulting in noncompliance with the U.S> Federal performance standard
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- Sedecal USA sent end user notification letters dated 9/26/02 to the consignees having the affected Sedecal Mobile X-Ray Systems, informing them that the units do not meet with the U.S. federal performance standards, that the labels are in Spanish and that the labels may not be visible after system installation. Sedecal schedule an appointment to conduct performance tests on the mobile x-ray systems to verify their compliance with the federal performance standard, submit the data to FDA, and place the English labels on the units. Sedecal will reimburse the owners for any out-of-pocket expenses incurred. Sedecal requested the accounts to call 800-920-9525 if they had any questions, and to inform Sedecal if they sold the unit, destroyed the unit or otherwise no longer had the unit in their possession.
- Quantity in commerce
- 9 units
- Distribution
- North Carolina, Louisiana, New York, New Jersey, Missouri, Kentucky, Texas and Germany
- Product codes and lots
- Model SM-20HF, serial numbers G13487, G13457; Model SM-30HF, serial numbers G12468, G13449, G13540, G13747, G13765, G14097, G14099, G14633, G15025