Z-0829-03
C1141 Flexible Endoscope Processing Container, designed to fit the Steris System I Processor. The unit may be...
STERIS Corporation / initiated 2003-04-23 / Terminated / root cause: sterility
Steris Corp began a recall of C1141 Flexible Endoscope Processing Container, designed to fit the Steris System I Processor. The unit may be sold individually or with the processing tray... on . The reason FDA records is "The container flow port is inverted, which restricts the flow of sterilant to the endoscope and affects the sterilization process". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-0829-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- STERIS Corporation
- Product
- C1141 Flexible Endoscope Processing Container, designed to fit the Steris System I Processor. The unit may be sold individually or with the processing tray assembly (C1140).
- Product code
- MLR Sterilizer, Chemical
- Reason for recall
- The container flow port is inverted, which restricts the flow of sterilant to the endoscope and affects the sterilization process.
- Root cause tag
- sterility
- FDA root cause
- Other
- Action
- The firm notified their customers by telephone and letter on 4/23/2003.
- Quantity in commerce
- 138 units
- Distribution
- The product was shipped to end-users located nationwide and worldwide.
- Product codes and lots
- All codes.