Z-0873-03

Dimension TSPA (Total Prostate Specific Antigen) Flex Reagent Cartridge. The product is shipped 4 cartridges...

Siemens Healthcare Diagnostics Inc. / initiated 2003-04-21 / Terminated

Dade Behring Inc began a recall of Dimension TSPA (Total Prostate Specific Antigen) Flex Reagent Cartridge. The product is shipped 4 cartridges per carton on . The reason FDA records is "test results may be inaccurate by 0.2 to 0.5 ng/mL". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0873-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension TSPA (Total Prostate Specific Antigen) Flex Reagent Cartridge. The product is shipped 4 cartridges per carton.
Product code
LTJ Prostate-Specific Antigen (Psa) For Management Of Prostate Cancers
Reason for recall
test results may be inaccurate by 0.2 to 0.5 ng/mL
FDA root cause
Other
Action
The recalling firm telephone each customer on 4/24-28/03 to determine if any customer used the lot (CD4085). If so, the customer was instructed to retest all samples with an alternate lot. All customers were instructed to discard any remaining inventory. As a follow up to the phone calls, the recalling firm issued a recall letter dated 4/25/03 informing the customer of the problem and to discard any remaining product.
Quantity in commerce
662 cartons
Distribution
The product was shipped to hospitals and clinic laboratories nationwide. There are eight government/military accounts; one of which is foreign.
Product codes and lots
Product code number RF451. Lot number CD4085 exp 3/25/04