Z-0873-03
Dimension TSPA (Total Prostate Specific Antigen) Flex Reagent Cartridge. The product is shipped 4 cartridges...
Siemens Healthcare Diagnostics Inc. / initiated 2003-04-21 / Terminated
Dade Behring Inc began a recall of Dimension TSPA (Total Prostate Specific Antigen) Flex Reagent Cartridge. The product is shipped 4 cartridges per carton on . The reason FDA records is "test results may be inaccurate by 0.2 to 0.5 ng/mL". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0873-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension TSPA (Total Prostate Specific Antigen) Flex Reagent Cartridge. The product is shipped 4 cartridges per carton.
- Product code
- LTJ Prostate-Specific Antigen (Psa) For Management Of Prostate Cancers
- Reason for recall
- test results may be inaccurate by 0.2 to 0.5 ng/mL
- FDA root cause
- Other
- Action
- The recalling firm telephone each customer on 4/24-28/03 to determine if any customer used the lot (CD4085). If so, the customer was instructed to retest all samples with an alternate lot. All customers were instructed to discard any remaining inventory. As a follow up to the phone calls, the recalling firm issued a recall letter dated 4/25/03 informing the customer of the problem and to discard any remaining product.
- Quantity in commerce
- 662 cartons
- Distribution
- The product was shipped to hospitals and clinic laboratories nationwide. There are eight government/military accounts; one of which is foreign.
- Product codes and lots
- Product code number RF451. Lot number CD4085 exp 3/25/04