Z-1088-03

Leonardo Workstations with software version 2022A/B, 2003A

Siemens Medical Solutions USA, Inc. / initiated 2003-02-28 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Leonardo Workstations with software version 2022A/B, 2003A on . The reason FDA records is "software problem This error may cause the loss of peripheral image information when there is a difference between the dimensions of the image matrix and the display segment". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1088-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Leonardo Workstations with software version 2022A/B, 2003A
Product code
LLZ System, Image Processing, Radiological
Reason for recall
software problem This error may cause the loss of peripheral image information when there is a difference between the dimensions of the image matrix and the display segment.
Root cause tag
software
FDA root cause
Other
Action
The recalling firm issued three separate customer advisory letters dated 2/28/03, 03/01/03, and 4/14/03 to their customers explaining the problem and the preventive measure to avoid loss of information.
Quantity in commerce
1374 units
Distribution
The product was shipped to distributors, medical facilities and Siemens training facilities nationwide.
Product codes and lots
Model 7129534, Serial # 01001 through 01380, 05001 through 05600, and 10001 through 10320