Z-0012-04

Roche/Hitachi Modular Analytics P clinical chemistry analyzer; Modular P. Catalog number 766-8156

Roche Diagnostics GmbH / initiated 2003-05-08 / Terminated

Roche Diagnostics Corp. began a recall of Roche/Hitachi Modular Analytics P clinical chemistry analyzer; Modular P. Catalog number 766-8156 on . The reason FDA records is "Incorrect results, without a data flag, reported if one of the other reagent containers becomes depleted before the R1 reagent is depleted". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0012-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Roche/Hitachi Modular Analytics P clinical chemistry analyzer; Modular P. Catalog number 766-8156.
Reason for recall
Incorrect results, without a data flag, reported if one of the other reagent containers becomes depleted before the R1 reagent is depleted.
FDA root cause
Other
Action
Customers were notified by certified mail on or about May 12, 2003. Operators were given a workaround (assure R1 has the lower number of tests remaining before using the analyzer) until new software is supplied.
Quantity in commerce
478
Distribution
United States
Product codes and lots
All units with current software versions (v 06-06 and 07-01).