Z-0012-04
Roche/Hitachi Modular Analytics P clinical chemistry analyzer; Modular P. Catalog number 766-8156
Roche Diagnostics GmbH / initiated 2003-05-08 / Terminated
Roche Diagnostics Corp. began a recall of Roche/Hitachi Modular Analytics P clinical chemistry analyzer; Modular P. Catalog number 766-8156 on . The reason FDA records is "Incorrect results, without a data flag, reported if one of the other reagent containers becomes depleted before the R1 reagent is depleted". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0012-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- Roche/Hitachi Modular Analytics P clinical chemistry analyzer; Modular P. Catalog number 766-8156.
- Reason for recall
- Incorrect results, without a data flag, reported if one of the other reagent containers becomes depleted before the R1 reagent is depleted.
- FDA root cause
- Other
- Action
- Customers were notified by certified mail on or about May 12, 2003. Operators were given a workaround (assure R1 has the lower number of tests remaining before using the analyzer) until new software is supplied.
- Quantity in commerce
- 478
- Distribution
- United States
- Product codes and lots
- All units with current software versions (v 06-06 and 07-01).