Z-0911-03

Clinitek 50 Urine Chemistry Analyzer; Product codes 6509C, 6513K, 6514C, 6515C, 6516F and 6518C. [The...

Medrad, Inc. / initiated 2003-06-19 / Terminated

Bayer HealthCare LLC, Diagnostics Div began a recall of Clinitek 50 Urine Chemistry Analyzer; Product codes 6509C, 6513K, 6514C, 6515C, 6516F and 6518C. [The different product codes are for the various countries... on . The reason FDA records is "Analyzer will produce erroneous results due to missing grounding strap". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0911-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medrad, Inc.
Product
Clinitek 50 Urine Chemistry Analyzer; Product codes 6509C, 6513K, 6514C, 6515C, 6516F and 6518C. [The different product codes are for the various countries where the product is distributed.]
Reason for recall
Analyzer will produce erroneous results due to missing grounding strap.
FDA root cause
Other
Action
Marketing sent a 'dear customer' letter to each account on May 13, 2003 telling them that their instrument had been identified as not meeting a production specification. An urgent product correction letter was subsequently sent to each account on or about June 20, 2003. The analyzers affected will be replaced.
Quantity in commerce
1324
Distribution
United States, Australia, Canada, Colombia, France, Germany, Hong Kong, India, Japan, Korea, Mexico, Sweden, Switzerland and United Kingdom.
Product codes and lots
serial numbers 168252 through 169574.