Z-0936-03
Abbott Latex-Free Primary I.V. Plumset Convertible Pin, 95 inch with Injection Site and Secure Lock, No...
Abbott Medical / initiated 2003-05-13 / Terminated
Abbott Laboratories HPD/ADD began a recall of Abbott Latex-Free Primary I.V. Plumset Convertible Pin, 95 inch with Injection Site and Secure Lock, No. 19020; Made in Costa Rica, Abbott Laboratories, North... on . The reason FDA records is "The I.V. administration set incorrectly lists the priming volume from the cassette inlet to distal male adapter as 13.5 mL instead of 11.3 mL". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0936-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Abbott Latex-Free Primary I.V. Plumset Convertible Pin, 95 inch with Injection Site and Secure Lock, No. 19020; Made in Costa Rica, Abbott Laboratories, North Chicago, IL 60064
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- The I.V. administration set incorrectly lists the priming volume from the cassette inlet to distal male adapter as 13.5 mL instead of 11.3 mL
- FDA root cause
- Other
- Action
- Recall letter dated 5/12/03 sent to the account on 5/13/03, informing them of the labeling discrepancy and requesting the return of any inventory of the affected lot to Abbott for credit.
- Quantity in commerce
- 4,656 units
- Distribution
- Miami, FL
- Product codes and lots
- List number 19020-04-12, lot 941845H