Z-0925-03

StatLock Tandem Microbore Catheter Extension Set with preattached Clearlink Valve 6'' LL. Model number 2N9205K

Venetec Intl., Inc. / initiated 2003-05-12 / Terminated

Venetec Intl Inc began a recall of StatLock Tandem Microbore Catheter Extension Set with preattached Clearlink Valve 6'' LL. Model number 2N9205K on . The reason FDA records is "Set tubing disconnected from the male luer lock during use on patients. Misassembled". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0925-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Venetec Intl., Inc.
Product
StatLock Tandem Microbore Catheter Extension Set with preattached Clearlink Valve 6'' LL. Model number 2N9205K.
Product code
FPA Set, Administration, Intravascular
Reason for recall
Set tubing disconnected from the male luer lock during use on patients. Misassembled.
FDA root cause
Other
Action
A recall letter was sent to the distributor who then sent thier own letters to the affected hospitals. Hospitals were reached on or about May 14, 2003 whereas the distributor was contacted by telephone and letter on or about May 12, 2003. It was verbally agreed that the distributor would subrecall the product.
Quantity in commerce
3150 units
Distribution
One distributor sent the product to 6 hospitals in the states of CT, IN, PA, TN.
Product codes and lots
Lot D207201