Z-0954-03
Evita 4 Continuous Ventilator
Draeger, Inc. / initiated 2003-05-07 / Terminated / root cause: software
Draeger Medical, Inc. began a recall of Evita 4 Continuous Ventilator on . The reason FDA records is "software problems - ventilation mode changes without user input, settings change without corresponding display change, false auto-set alarm". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0954-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Draeger, Inc.
- Product
- Evita 4 Continuous Ventilator
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- software problems - ventilation mode changes without user input, settings change without corresponding display change, false auto-set alarm.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- The recalling firm issued recall letters to the accounts dated 5/7/03.
- Quantity in commerce
- 806 units
- Distribution
- The product was shipped nationwide to distributor, hospitals and a physician. The product was also shipped to medical facilities and universities in Canada.
- Product codes and lots
- Catalog numbers 4116640 and 8412980