Z-1130-03
Free T4 Control Pack, list 9C03-10; the kit consists of 3 bottles of controls, L - 0.7, M - 1.2 and H - 3.0...
Abbott Medical / initiated 2003-05-30 / Terminated / root cause: labeling and UDI
Abbott Laboratories HPD/ADD began a recall of Free T4 Control Pack, list 9C03-10; the kit consists of 3 bottles of controls, L - 0.7, M - 1.2 and H - 3.0; Abbott Laboratories, Abbott Park, IL 60064 on . The reason FDA records is "The Free T4 Controls, may yield control values that are high and outside of the package insert ranges". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1130-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Free T4 Control Pack, list 9C03-10; the kit consists of 3 bottles of controls, L - 0.7, M - 1.2 and H - 3.0; Abbott Laboratories, Abbott Park, IL 60064
- Product code
- CEC Radioimmunoassay, Free Thyroxine
- Reason for recall
- The Free T4 Controls, may yield control values that are high and outside of the package insert ranges.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- The Free T4 Control Pack, list 9C03-10, and the IMx Free T4 Reagent Kit, list 2222-20, were recalled by letter dated 5/30/03. The accounts were informed of the out of range control values, and the potential for depressed patient values with the IMx kit, and were requested to discontinue use and destroy the affected lots of product. Abbott sent a Device Correction letter dated 6/4/03 to those accounts receiving the Architect Free T4 Calibrator kits, informing them of the problems experienced with the The Free T4 Control Pack, list 9C03-10, and the IMx Free T4 Reagent Kit, list 2222-20, made from the same raw material. The accounts were informed that the Architect Calibrator was performing as expected, but as a precaution recommended conducting a daily QC with non-Abbott Controls and to report back to Abbott any out of range non-Abbott control results.
- Quantity in commerce
- 3247 kits
- Distribution
- Nationwide and internationally to Abbott subsidiaries in Ecuador, Uruguay, Argentina, Brazil, El Salvador, Canada, Germany, Japan, Singapore, Hong Kong, Thailand, Korea, Australia, Costa Rica, Grenada, St. Lucia, Bermuda, Antigua, Panama, Guatemala, Mexico, Colombia, Jamaica, Paraguay, Peru, Chile, Dominican Republic, United Kingdom, New Zealand, Trinidad & Tabago, St. Vincent, Barbados, Curacao, St. Kitts, Anguilla, Turks & Caicos Islands, Cayman Islands, Germany and Taiwan.
- Product codes and lots
- List 9C03-10, lot numbers 95881Q100 and 95881Q101