Z-0968-03
Bulk 16 FR Valved PTFE Sheath/Dilator, Part Number 10614-011
Greatbatch Medical / initiated 2003-06-06 / Terminated
MedAmicus,Inc began a recall of Bulk 16 FR Valved PTFE Sheath/Dilator, Part Number 10614-011 on . The reason FDA records is "Stability testing, after accelerated aging of the device, showed that the handle of device did not meet strength specifications". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0968-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Greatbatch Medical
- Product
- Bulk 16 FR Valved PTFE Sheath/Dilator, Part Number 10614-011
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Stability testing, after accelerated aging of the device, showed that the handle of device did not meet strength specifications.
- FDA root cause
- Other
- Action
- The consignees were sent recall letters, dated June 6, 2003 or June 20, 2003, that requested the return of the products.
- Distribution
- The products were distributed to 4 consignees in Georgia, Missouri, New York and North Carolina. Most of the product was shipped to a distributor in New York which probably shipped to other states. There was no direct distribution to foreign or U.S Government accounts.
- Product codes and lots
- Lot Number W24540