75-1650 Drip Chambers .138rr; Abbott, Ashland, Ohio 44805; 800 drip chambers per case; Abbott Commodity...
Abbott Laboratories HPD/ADD began a recall of 75-1650 Drip Chambers .138rr; Abbott, Ashland, Ohio 44805; 800 drip chambers per case; Abbott Commodity 751650, Ashland Product #85-16500 on . The reason FDA records is "The drip chambers and/or adapters in the sets were manufactured with non-radiation grade plastic resin, resulting in discoloration of the resin". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1136-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- 75-1650 Drip Chambers .138rr; Abbott, Ashland, Ohio 44805; 800 drip chambers per case; Abbott Commodity 751650, Ashland Product #85-16500
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- The drip chambers and/or adapters in the sets were manufactured with non-radiation grade plastic resin, resulting in discoloration of the resin.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- The IV administration sets were recalled by letters dated 6/9/03, advising the accounts of the discoloration of the sight chamber and adapters due to the use of non-radiation grade plastic resin in the manufacture of these components of the sets and requesting the return of the sets for credit. These letters were accompanied by a customer ship history/business reply form, which listed the specific set list numbers and lot numbers shipped to the direct account. The direct accounts were requested to send copies of the enclosed recall letter with the listing of all affected sets to their subaccounts. Abbott sent recall letters dated 6/13/03 to those accounts who received I.V. set list 018810458, lot 910824W packed in cases with Liposyn II 20% Kits, list 97920403, lot 92552DF and list 97930403, lot 94553DF. These accounts were informed of the problem with the administration sets and were requested to remove the sets from the Liposyn II cases and return the sets for credit. Abbott sent recall letters dated 7/7/03 to distributors and veterinary clinics receiving the 12 additional lots of primary I.V. sets, list numbers 018810448 and 089620448, informing them of the problem with the drip chambers and requesting them to return the affected lots for credit. The distributors were requested to notify their customers. Abbott sent recall letters dated 7/7/03 to the third party manufacturers receiving the affected drip chamber components, informing them of the problem with the drip chambers and requesting the return of the affected lots. Abbott sent recall letters dated 8/19/03 to hospitals and distributors receiving the 9 additional lots of I.V. sets, list numbers 119440402, 119480402, 120300412, 122590402 and 123120401, informing them of the problem with the drip chambers and requesting them to return the affected lots for credit. The distributors were requested to notify their customers. Abbott sent recall letters dated 5/13/04 to hospitals and distributor
- Quantity in commerce
- 58,400 drip chambers
- Distribution
- Nationwide and Internationally to Abbott affiliates in Argentina, Aruba, Australia, the Bahamas, Barbados, Bermuda, Bolivia, Brazil, Canada, Cayman Islands, Chile, Columbia, Costa Rica, Croatia, Curacao, Dominican Republic, Ecuador, Egypt, El Salvador, France, Greece, Guam, Guatemala, Haiti, Honduras, Hong Kong, Hungary, Italy, Jordan, Kuwait, Mexico, New Zealand, Nicaragua, Pakistan, Panama, Paraguay, Peru, Suadia Arabia, Singapore, Taiwan, Trinidad, Turkey, United Arab Emirates and Uruguay
- Product codes and lots
- Abbott Commodity 751650, Ashland Product #85-16500, lots 85216NLMP, 85464NLMP, 85651NLMP, 85696NLMP, 85785NLMP