Z-1061-03
ASAP Channel Cut Biopsy System, 14 gauge Reference Number: 1250
Boston Scientific Corporation / initiated 2003-07-03 / Terminated / root cause: packaging
Boston Scientific Corporation began a recall of ASAP Channel Cut Biopsy System, 14 gauge Reference Number: 1250 on . The reason FDA records is "Sterility of the device may be compromised due to a lack of package integrity". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1061-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- ASAP Channel Cut Biopsy System, 14 gauge Reference Number: 1250
- Product code
- KNW Instrument, Biopsy
- Reason for recall
- Sterility of the device may be compromised due to a lack of package integrity
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Boston Scientific notified accounts by letter on 7/3/03. Accounts are requested to check inventory and return product. A Veification Form was requested to be completed.
- Quantity in commerce
- 1,030 units
- Distribution
- Nationwide
- Product codes and lots
- 4104836, 4145036, 4211323, 4290090 ,4370503 4405122, 4445839, 4461241, 4569317, 4666229 4916051, 5022356, 5120746, 5133248, 5251508, 5288530