Z-1064-03
Servo Controlled Oxygen System for Ohmeda''s Medical Giraffe OmniBed and Ohmeda''s Medical Girraffe Incubators
GE Healthcare / initiated 2003-06-13 / Terminated / root cause: manufacturing process
Ohmeda Medical, A division of Datex-Ohmeda, Inc. began a recall of Servo Controlled Oxygen System for Ohmeda''s Medical Giraffe OmniBed and Ohmeda''s Medical Girraffe Incubators on . The reason FDA records is "Firm's received complaints of medical device malfunction affecting instrument's calibration and alarm systems". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1064-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Servo Controlled Oxygen System for Ohmeda''s Medical Giraffe OmniBed and Ohmeda''s Medical Girraffe Incubators
- Product code
- 80FMZ
- Reason for recall
- Firm's received complaints of medical device malfunction affecting instrument's calibration and alarm systems.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- Recalling firm notified foreign distributors and domestic consignees by e-mail and phone on 6/13/2003. Foreign regional distributors were requested to manage the recall. The notification described various instrument problems and remedial action available via the firm''s field action retrofit kits, #6600-0126-850 for the Giraffe OmniBed and #6600-0261-850 for the Giraffe Incubator.
- Quantity in commerce
- 157 units
- Distribution
- Devices were sold to 15 international distributors and one domestic end user.
- Product codes and lots
- Giraffe OmniBeds (w/Servo O2 option): HEDF50041 - HEDF50076 HDGE50562 ¿ HDGE50567, HDGF54039, HDGF54153 ¿ HDGF54155 HDGF54165 ¿ HDGF54167 Giraffe Incubators (w/Servo O2 option): HDHE50160-HDHE50161, HDHF50023-HDHF50024, HDHF50105, HDHF50109, HDHF50114, HDHF50227- HDHF50245, HDHF50278, HDHF50302, HDHF50304, HDHF50306, HDHF50309, HDHF50327-HDHF50328, HDHF50333-HDHF50343, HDHF53070, HDHF54046, HDHF54122, HDHF54131-HDHF54133, HDHF54135, HDHF54143-HDHF54145, HDHF54152-HDHF54155.