Z-1161-03

Clearlink System Y-Type Blood/Solution Set, 10 drops/mL, 112'', Standard Blood Filter, 170 to 260 Micron...

Baxter Healthcare Corporation / initiated 2003-07-16 / Terminated / root cause: packaging

Baxter Healthcare Corp. began a recall of Clearlink System Y-Type Blood/Solution Set, 10 drops/mL, 112'', Standard Blood Filter, 170 to 260 Micron Filter, Luer Activated Valve, Male Luer Lock Adapter... on . The reason FDA records is "There are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1161-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Baxter Healthcare Corporation
Product
Clearlink System Y-Type Blood/Solution Set, 10 drops/mL, 112'', Standard Blood Filter, 170 to 260 Micron Filter, Luer Activated Valve, Male Luer Lock Adapter, catalog 2C8750; Baxter Healthcare Corporation, Deerfield, IL 60015 USA
Product code
KSB Set, Transfer (Blood/Plasma)
Reason for recall
There are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process.
Root cause tag
packaging
FDA root cause
Other
Action
Recalled by letter dated 7/16/03. The accounts were informed of the potential for leaks in the set tubing, and asked to inspect their inventory for the affected product codes and lots, segregate any affected products found and discard them according to local procedures. The customers will receive credit for the discarded affected product.
Quantity in commerce
166,320 units
Distribution
Nationwide, Mexico, Canada, Columbia, Australia, New Zealand and Taiwan
Product codes and lots
catalog 2C8750, lot numbers UR128876 through UR168104. Please note: Baxter lot numbers are represented by an internal reference code ''UR'' followed by a six digit sequential number, or by the letter ''U'' followed by a six digit sequential number with and ''R'' suffix. Any lot number appearing within the identified six digit sequential number range is subject to the recall.