Z-1163-03
Fenwal Y-Type Blood/Solution Set with Large Standard Blood Filter, 15 drops/mL, 89'', 170 to 260 Micron...
Baxter Healthcare Corporation / initiated 2003-07-16 / Terminated / root cause: packaging
Baxter Healthcare Corp. began a recall of Fenwal Y-Type Blood/Solution Set with Large Standard Blood Filter, 15 drops/mL, 89'', 170 to 260 Micron Filter, Large Bore Tubing, Male Luer Lock Adapter... on . The reason FDA records is "There are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1163-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Fenwal Y-Type Blood/Solution Set with Large Standard Blood Filter, 15 drops/mL, 89'', 170 to 260 Micron Filter, Large Bore Tubing, Male Luer Lock Adapter, catalog 4C7776A; Baxter Healthcare Corporation, Deerfield, IL 60015 USA
- Product code
- BRZ Set, Blood Transfusion
- Reason for recall
- There are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Recalled by letter dated 7/16/03. The accounts were informed of the potential for leaks in the set tubing, and asked to inspect their inventory for the affected product codes and lots, segregate any affected products found and discard them according to local procedures. The customers will receive credit for the discarded affected product.
- Quantity in commerce
- 21,024 units
- Distribution
- Nationwide, Mexico, Canada, Columbia, Australia, New Zealand and Taiwan
- Product codes and lots
- catalog 4C7776A, lot numbers UR160143 through UR166538. Please note: Baxter lot numbers are represented by an internal reference code ''UR'' followed by a six digit sequential number, or by the letter ''U'' followed by a six digit sequential number with and ''R'' suffix. Any lot number appearing within the identified six digit sequential number range is subject to the recall.