Z-1144-03

Bardex Foley Catheter, 22 Fr 40cc- 20cc Ribbed Balloons, Coleman Model Red Latex, 3 Way Radiopaque, Lubricath...

C.R. Bard Inc / initiated 2002-08-29 / Terminated / root cause: labeling and UDI

C.R. Bard, Inc., Urological Division began a recall of Bardex Foley Catheter, 22 Fr 40cc- 20cc Ribbed Balloons, Coleman Model Red Latex, 3 Way Radiopaque, Lubricath Coated, Sterile, Single Use Device, C.R. Bard... on . The reason FDA records is "On some units the inflation valve labeling is reversed on the Prostate and Bladder balloon inflation ports". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1144-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
C.R. Bard Inc
Product
Bardex Foley Catheter, 22 Fr 40cc- 20cc Ribbed Balloons, Coleman Model Red Latex, 3 Way Radiopaque, Lubricath Coated, Sterile, Single Use Device, C.R. Bard, Inc., Covington, GA 30014. The catheter is an open whistle tip with two eyes on the same side. It is made of red latex for added catheter stiffness and radiopacity, with a Lubricath hydrophilic coating. This product has 3 lumens; a lumen for drainage, a lumen for inflation of the bladder balloon, and a lumen for inflation of the prostatic balloon.
Product code
MJC Catheter, Urological (Antimicrobial) And Accessories
Reason for recall
On some units the inflation valve labeling is reversed on the Prostate and Bladder balloon inflation ports.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
Consignees were notified by letter on 8/29/2002. They were advised to pretest the catheter prior to inserting the catheter into the urethra.If the pretest indicates that the catheter balloons do not inflate as labeled, they were advised not to use the product and to contact Bard Medical Division Field Assurance Manager to have the product returned and credited.
Quantity in commerce
1461 units
Distribution
The product was distributed nationwide, Canada, Belgium and Germany.
Product codes and lots
Catalog Number 6016L22 (all units packaged from January 1999 and shipped prior to September 2002).