Z-0051-04

propper brand cal-mark(R), Micro-Hematocrit Calibrated Capillary Tubes, CALIBRATED-HEPARINIZED, reorder no...

Propper Mfg Co Inc / initiated 2002-08-09 / Terminated

Propper Mfg Co Inc began a recall of propper brand cal-mark(R), Micro-Hematocrit Calibrated Capillary Tubes, CALIBRATED-HEPARINIZED, reorder no. 043021, 6 Vials, 100 Tubes per Vial, Length 75 mm... on . The reason FDA records is "Deviation in capillary tubes could cause a 10% increase in read out values which could make a slight anemic patient appear to have a normal reading". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0051-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Propper Mfg Co Inc
Product
propper brand cal-mark(R), Micro-Hematocrit Calibrated Capillary Tubes, CALIBRATED-HEPARINIZED, reorder no. 043021, 6 Vials, 100 Tubes per Vial, Length 75 mm., 0.55 mm. I.D., 1.40-1.75mm. O.D. Calibrated 60mm from end for use in direct reading centrifuges. Fill to the 60mm mark with blood and seal opposite end by flame or with sealing compound. If centrifuged on a direct reading centrifuge, read percent hematocrit directly from scale on the centrifuge. Otherwise, read in normal manner as a percent ratio of packed cells with 60mm of sample as 100%. Firm on label: propper manufacturing co., inc., long island city, n.y. 11101.
Product code
G-IO
Reason for recall
Deviation in capillary tubes could cause a 10% increase in read out values which could make a slight anemic patient appear to have a normal reading.
FDA root cause
Other
Action
On 8/9/02, Propper faxed a 'Request for Return' of the product to the consignees advising them of the problem and that a replacement will be sent for all returned product. On 11/27/02, Propper faxed a second 'Request for Return of Product' to its consignees. The firm destroyed the recovered and held stock.
Quantity in commerce
178 boxes
Distribution
Distributors located in TX, UT, ID, IL, IA, IN, MN, WI, OH, CT, MO, PA, VA.
Product codes and lots
Reorder No. 043021, Lot # C0402