Z-1122-03
Radionics Cool-Tip RF Electrode Kit, containing DGP-HP single-use grounding pads
Medtronic, Inc. / initiated 2003-07-16 / Terminated
Valleylab, Inc began a recall of Radionics Cool-Tip RF Electrode Kit, containing DGP-HP single-use grounding pads on . The reason FDA records is "Pads may have a thin plastic film covering the conductive gel surface underneath the normal printed cover, causing non-adherence and possible burn". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1122-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Radionics Cool-Tip RF Electrode Kit, containing DGP-HP single-use grounding pads.
- Product code
- HAM Apparatus, Electrosurgical
- Reason for recall
- Pads may have a thin plastic film covering the conductive gel surface underneath the normal printed cover, causing non-adherence and possible burn.
- FDA root cause
- Other
- Action
- Users were notified by letter on 7/16/2003.
- Quantity in commerce
- 10,000 kits
- Distribution
- Nationwide. VA Hospital, San Antonio, TX. Foreign distribution to Australia, Brazil, Canada, Dominican Republic, France, Germany, Hong Kong, India, Israel, Japan, Korea, Phillipines, Singapore, Taiwan.
- Product codes and lots
- Catalog Numbers: CT-1020, CT-1030, CT-1510, CT-1520, CT-1530, CT-2020, CT-2030, CT-2530, CTC-1025, CTC-1525, CTC-2025.