Z-1071-03
Coulter Clone B6-FITC Reagent
Siemens Healthcare Diagnostics Inc. / initiated 2003-04-28 / Terminated
Beckman Coulter Inc began a recall of Coulter Clone B6-FITC Reagent on . The reason FDA records is "Diminished expression on B-Cell populations when drawn in EDTA tubes, which may lead to inaccurate interpretation of phenotype results. It was determined that choice of anticoagulant may have an effect on results for...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1071-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Coulter Clone B6-FITC Reagent
- Reason for recall
- Diminished expression on B-Cell populations when drawn in EDTA tubes, which may lead to inaccurate interpretation of phenotype results. It was determined that choice of anticoagulant may have an effect on results for this reagent. The CD23 clone has shown a diminished expression on B-Cell populations when drawn on EDTA tubes, which may lead to inaccurate interpretation of phenotype results.
- FDA root cause
- Other
- Action
- Customer letters will be sent to all known purchasers on or about 4/28/2003, also existing inventory will be reworked to include a letter describing performance issues when collection occurs using EDTA.
- Quantity in commerce
- 122
- Distribution
- Nationwide and to Canada
- Product codes and lots
- Lots 735512F, 735513F and 735514FR