Immulite 2000. Immunoassay system. Software version 2.6 was to be uploaded on systems running 2.5b software

Siemens Healthcare Diagnostics Inc. / initiated 2003-06-13 / Terminated / root cause: software

DPC Cirrus began a recall of Immulite 2000. Immunoassay system. Software version 2.6 was to be uploaded on systems running 2.5b software on . The reason FDA records is "Version 2.6 software was released however it will not properly handle adjustment slopes". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1138-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Immulite 2000. Immunoassay system. Software version 2.6 was to be uploaded on systems running 2.5b software.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Version 2.6 software was released however it will not properly handle adjustment slopes.
Root cause tag
software
FDA root cause
Other
Action
Recall Technical Bulletins were sent out on 6/13/2003 to all Immulite 2000 users instructing them to not install the provided version 2.6 software.
Quantity in commerce
434 upgrades
Distribution
Software was distributed to Immulite 2000 users throughout the nation. These units are found in hospitals and clinics. Software was also distributed to the firm''s service representatives.
Product codes and lots
All units.