Z-1138-03
Immulite 2000. Immunoassay system. Software version 2.6 was to be uploaded on systems running 2.5b software
Siemens Healthcare Diagnostics Inc. / initiated 2003-06-13 / Terminated / root cause: software
DPC Cirrus began a recall of Immulite 2000. Immunoassay system. Software version 2.6 was to be uploaded on systems running 2.5b software on . The reason FDA records is "Version 2.6 software was released however it will not properly handle adjustment slopes". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1138-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Immulite 2000. Immunoassay system. Software version 2.6 was to be uploaded on systems running 2.5b software.
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Version 2.6 software was released however it will not properly handle adjustment slopes.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- Recall Technical Bulletins were sent out on 6/13/2003 to all Immulite 2000 users instructing them to not install the provided version 2.6 software.
- Quantity in commerce
- 434 upgrades
- Distribution
- Software was distributed to Immulite 2000 users throughout the nation. These units are found in hospitals and clinics. Software was also distributed to the firm''s service representatives.
- Product codes and lots
- All units.