Z-1208-03
Software Versions V1.2R,or earlier for the Harvard 2, Dual Infusion Pump (P/N 2002-001)
Harvard Clinical Technology / initiated 2003-08-11 / Terminated
Harvard Clinical Technology began a recall of Software Versions V1.2R,or earlier for the Harvard 2, Dual Infusion Pump (P/N 2002-001) on . The reason FDA records is "Under specific sequence conditions, the unit may apper to be infusing, but is not". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1208-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Harvard Clinical Technology
- Product
- Software Versions V1.2R,or earlier for the Harvard 2, Dual Infusion Pump (P/N 2002-001)
- Product code
- FRN Pump, Infusion
- Reason for recall
- Under specific sequence conditions, the unit may apper to be infusing, but is not
- FDA root cause
- Other
- Action
- Harvard Medical Technology inc. notified customers by letter on 8/11/03, advising users of the software upgrade to version V1.3T. Facilities will be visited by technicians who will upgrade the software.
- Quantity in commerce
- 384 units
- Distribution
- Nationwide
- Product codes and lots
- Software versions V1.2R or earlier