Z-1186-03
Alumina ceramic liner for acetabular shell. This is part of the Transcend ceramic-on-ceramic total hip system
Wright Medical Technology Inc / initiated 2003-08-18 / Terminated
Wright Medical Technology Inc began a recall of Alumina ceramic liner for acetabular shell. This is part of the Transcend ceramic-on-ceramic total hip system on . The reason FDA records is "Wright Medical commercially distributed Transcend alumina acetabular shell liners lacking the required process validation". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1186-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Wright Medical Technology Inc
- Product
- Alumina ceramic liner for acetabular shell. This is part of the Transcend ceramic-on-ceramic total hip system.
- Product code
- MRA Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented
- Reason for recall
- Wright Medical commercially distributed Transcend alumina acetabular shell liners lacking the required process validation.
- FDA root cause
- Other
- Action
- Distributors returned thier stock after May 8, 2003
- Quantity in commerce
- 133 units
- Distribution
- nationwide
- Product codes and lots
- Part No. Lot Serial No. 73003258 070A122950 71002846 101A147607 72003252 101A147579 71002846 01214750 73003258 020A116075 73003258 101A147616 71002846 101A147609 71002846 01214749 72003252 01214581 72003252 01215268 73003258 101A147617 73003258 01214576 74003664 101A147618 72003252 101A147580 71002846 048A080430 72003252 0124580 71002846 01214751 74003664 061A136523 72003252 119A112057 73003258 01214575 71002846 031A139399 72003252 101A147577 72003252 01214582 72003252 101A147578 71002846 039A100609 72003252 081A145884