Vitros Chemistry *** GLU (glucose) Slides *** 5 cartridges each containing 60 slides for a total of 300 tests...
Ortho-Clinical Diagnostics began a recall of Vitros Chemistry *** GLU (glucose) Slides *** 5 cartridges each containing 60 slides for a total of 300 tests per box *** REF 170 7801 *** Ortho-Clinical... on . The reason FDA records is "Results of glucose measurements on sodium fluoride/potassium oxalate specimens can be negatively biased when using Vitros GLU and GLU DT slides stored in the refrigerator. GLU slides are processed by high volume...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1224-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- Vitros Chemistry *** GLU (glucose) Slides *** 5 cartridges each containing 60 slides for a total of 300 tests per box *** REF 170 7801 *** Ortho-Clinical Diagnostics Inc., 100 Indigo Creek Drive, Rochester, NY 14626. [NOTE: product has an 18 month shelf life.]
- Product code
- CGA Glucose Oxidase, Glucose
- Reason for recall
- Results of glucose measurements on sodium fluoride/potassium oxalate specimens can be negatively biased when using Vitros GLU and GLU DT slides stored in the refrigerator. GLU slides are processed by high volume chemistry systems typically used in hospital laboratories.
- FDA root cause
- Other
- Action
- GLU Slides: Letters dated between 7/28/03 and 8/11/03 provide revised storage instructions. Customers using NaF/KOx blood collection tubes and have VITROS GLU slides that have been refrigerated for more than four months will receive replacement product. Those customers are instructed to continue to use the slides only until replacement product arrives. GLU DT Slides: Letters dated between 7/28/03 and 8/11/03 with the same instructions as for GLU slides.
- Quantity in commerce
- 487,912 boxes
- Distribution
- Nationwide and worldwide distribution to distributors and end-users.
- Product codes and lots
- All lots with an expiration date of 1 December 2004 and earlier.