Vitros Chemistry *** GLU (glucose) DT Slides ***25 slides per box *** CAT 153 2316 *** Ortho-Clinical...
Ortho-Clinical Diagnostics began a recall of Vitros Chemistry *** GLU (glucose) DT Slides ***25 slides per box *** CAT 153 2316 *** Ortho-Clinical Diagnostics Inc. , 100 Indigo Creek Dr., Rochester, NY... on . The reason FDA records is "Results of glucose measurements on sodium fluoride/potassium oxalate specimens can be negatively biased when using Vitros GLU and GLU DT slides stored in the refrigerator". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1225-03
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- Vitros Chemistry *** GLU (glucose) DT Slides ***25 slides per box *** CAT 153 2316 *** Ortho-Clinical Diagnostics Inc. , 100 Indigo Creek Dr., Rochester, NY 14626. [NOTE: product has a 24-month shelf life.]
- Product code
- CGA Glucose Oxidase, Glucose
- Reason for recall
- Results of glucose measurements on sodium fluoride/potassium oxalate specimens can be negatively biased when using Vitros GLU and GLU DT slides stored in the refrigerator.
- FDA root cause
- Other
- Action
- GLU Slides: Letters dated between 7/28/03 and 8/11/03 provide revised storage instructions. Customers using NaF/KOx blood collection tubes and have VITROS GLU slides that have been refrigerated for more than four months will receive replacement product. Those customers are instructed to continue to use the slides only until replacement product arrives. GLU DT Slides: Letters dated between 7/28/03 and 8/11/03 with the same instructions as for GLU slides.
- Quantity in commerce
- 210,050 boxes
- Distribution
- Nationwide and worldwide distribution to distributors and end-users.
- Product codes and lots
- All lots with an expiration date of 1June 2005 and earlier.