Z-0053-04
e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens...
Siemens Medical Solutions USA, Inc. / initiated 2003-09-05 / Terminated
Siemens Medical Solutions USA, Inc. began a recall of e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens Medical Systems, Inc., Nuclear Medicine Group... on . The reason FDA records is "The detector's radial brake may not be applied properly, allowing unintended motion which could result in injury to the patient or operator". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0053-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL 60195-5203
- Product code
- KPS--
- Reason for recall
- The detector's radial brake may not be applied properly, allowing unintended motion which could result in injury to the patient or operator.
- FDA root cause
- Other
- Action
- Siemens sent Important Safety Customer Advisory letters dated 8/27/03 to all e.cam customers on 9/5-12/03. The accounts were informed of the problem with the detector''s radial brake not being applied properly, allowing slow, unintended downward motion of the detector that could result in injury to the patient or operator, and were advised not to leave the patient unattended while on the pallet and not to leave the patient under the detector upon completion of the study. The customers were advised that they would be contacted by their Siemens service provider to schedule a mandatory update to the e.cam to correct the problem. The upgrade will be done free of charge.
- Quantity in commerce
- 2,386 units
- Distribution
- Nationwide, United Arab Emirates, Albania, Argentina, Austria, Australia, Bosnia, Barbados, Bangladesh, Belgium, Bolivia, Brazil, Cameroon, Canada, Chile, China, Colombia, Republic of Congo, Costa Rica, Czech Republic, Denmark, Dominican Republic, El Salvador, Ecuador, Egypt, France, Finland, Germany, Great Britain, Georgia, Greece, Hong Kong, Honduras, Ireland, India, Italy, Iran, Iraq, Jordan, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Mexico, Morocco, Macedonia, Madagascar, Myanmar, Mauritius, Malaysia, Nepal, Norway, New Zealand, Oman, Peru, Philippines, Pakistan, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, Ukraine, Uzbekistan, Zimbabwe, South Africa and the Netherlands.
- Product codes and lots
- all e.cam models: 4380213, 4380221, 5242826, 5977066, 5977074, 5984005, 5989079, 5989087, 5989095, 5991109, 5991117, 5992099, 7332625, 7332633, 7760809, 7760932, 7761161, 7823920, 7823938, 7823946, 7823953, 7823979; serial numbers 1 through 9091