Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0023-04

HYDRO-PRO brand Suction-Irrigation System without Probe, Disposable (single use only), sterile, latex free...

Vital Concepts, Inc. / initiated 2003-07-11 / Terminated

Vital Concepts, Inc. began a recall of HYDRO-PRO brand Suction-Irrigation System without Probe, Disposable (single use only), sterile, latex free. R-Med Re-Order # 55170R; Distributed by R-Med... on . The reason FDA records is "The suction valve may become stuck in the 'suction on' position under vacuum". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0023-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Vital Concepts, Inc.
Product
HYDRO-PRO brand Suction-Irrigation System without Probe, Disposable (single use only), sterile, latex free. R-Med Re-Order # 55170R; Distributed by R-Med, Inc., Orgeon, OH
Product code
GCY Apparatus, Suction, Single Patient Use, Portable, Nonpowered
Reason for recall
The suction valve may become stuck in the 'suction on' position under vacuum.
FDA root cause
Other
Action
A request for product return was sent to each customer beginning on July 11, 2003.
Quantity in commerce
288
Distribution
Arizona, Illinois, Indiana, Michigan, Ohio, Pennsylvania, South Carolina, Texas, Utah, Washington, Canada, Israel and Japan
Product codes and lots
lots 50904954, 50905019 and 50905099.