Z-0564-04
CordisEndomyocardial Biopsy Forceps 7F 104 cm Standard, Product Number 504300L
Cordis US Corp. / initiated 2003-08-29 / Terminated / root cause: labeling and UDI
Cordis Corporation began a recall of CordisEndomyocardial Biopsy Forceps 7F 104 cm Standard, Product Number 504300L on . The reason FDA records is "Some devices may have an incorrect inner label which shows Product # 504302L with a length of 50cm, while the outer label is correctly labeled as Product # 504300L with the length of 104mm". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0564-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cordis US Corp.
- Product
- CordisEndomyocardial Biopsy Forceps 7F 104 cm Standard, Product Number 504300L
- Product code
- GEN Forceps, General & Plastic Surgery
- Reason for recall
- Some devices may have an incorrect inner label which shows Product # 504302L with a length of 50cm, while the outer label is correctly labeled as Product # 504300L with the length of 104mm.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- The firm initially sent Cordis representatives to visit all 21 domestic accounts the week of August 25, 2003. Follow-up Recall notificaiton Letters were sent to each account discussing the problem and informing them that a Cordis Representative has been at their location and inspected their product for misbranded devices. In 18 cases the letter stated that NO mislabeled product was found but in three other cases they were told that mislabeled product was found and recovered. These follow-up letters to the visits were sent out on August 29, 2003. A similar inspection was performed by Cordis Representatives for International Accounts and the Recall Notification was sent to the International Accounts on 8/29/2003.
- Quantity in commerce
- 220
- Distribution
- The devices were distributed to 21 domestic hospital accounts in MI, OK, FL, TN, NY, IN, VA, NC, WI, GA, IA, NE, CA. Also product was distributed to International Distributors in the Netherlands, Canada and Mexico.
- Product codes and lots
- Lot Number 70403255