Lactate Disposable Membrane Caps Product Number: 0018108400

Instrumentation Laboratory Co. / initiated 2003-08-28 / Terminated

Instrumentation Laboratory Co. began a recall of Lactate Disposable Membrane Caps Product Number: 0018108400 on . The reason FDA records is "Lactate Disposable Membrane Caps linearity was determined to be significantly biased and could potentially cause erroneous patient results above 6.0 mmol/L". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0027-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Instrumentation Laboratory Co.
Product
Lactate Disposable Membrane Caps Product Number: 0018108400
Product code
LNW Paper, Obstetric Ph
Reason for recall
Lactate Disposable Membrane Caps linearity was determined to be significantly biased and could potentially cause erroneous patient results above 6.0 mmol/L.
FDA root cause
Other
Action
Instrumentation Laboratories notified two (2) U.S. direct customers on August 28, 2003 by telephone, regarding Lot N30601. Follow-up phoning emphasizing significance of linearity problem (and additional involved lot N307 - for 1 affected customer) performed on September 3, 2003. Phone call followed with faxback acknowledging notification and disposition of remaining product lot(s). IL service also dispatched to customer sites on September 4, 2003. International Distributor Notices sent to IL Europe and IL Mexico on September 4, 2003. Product Information Bulletins circulated on September 3, 2003, to IL personnel worldwide.
Quantity in commerce
287 boxes
Distribution
IL, NY International: Italy, Mexico
Product codes and lots
US Distribution Lot Numbers: N307, Exp. 1/30/04  N30601, Exp 12/30/03 International Distribution: N30301 Exp 9/30/03 N30501 EXp 11/30/03 N306 Exp 12/30/03 N30604 Exp 12/30/03 N308 Exp 2/28/04