Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1259-03

ETHICON brand surgical device with an added label with the labeling Manufactured by: MedSurg Solutions, 3025...

International Surgical Supply, Inc. / initiated 2003-09-03 / Terminated / root cause: sterility

International Surgical Supply, Inc. began a recall of ETHICON brand surgical device with an added label with the labeling Manufactured by: MedSurg Solutions, 3025 Harbor Lane suite 104 Plymouth, MN 55447 and/or... on . The reason FDA records is "Various Sterile Expired and Non-expired Ethicon brand surgical devices were resterilized without adequate validations and distributed". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-1259-03
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
International Surgical Supply, Inc.
Product
ETHICON brand surgical device with an added label with the labeling Manufactured by: MedSurg Solutions, 3025 Harbor Lane suite 104 Plymouth, MN 55447 and/or another additional label which states MEDSURG 3025 Harbor Lane, Suite 104, Plymouth, MN 55447 Steriliation Date:. The product labels also state the following firm: ETHICON, INC. SOMERVILLE, N.J. 08876-0151 The product described on label is a: Reloads for Devices EZ45, ET45, NK45 Thick Tissue, item ZR45G
Product code
KDC Instrument, Surgical, Disposable
Reason for recall
Various Sterile Expired and Non-expired Ethicon brand surgical devices were resterilized without adequate validations and distributed.
Root cause tag
sterility
FDA root cause
Other
Action
The firm sent an e-mail recall notification and request for response on 9/3/2003, to the one direct account in Minnesota requesting quarantine and return of product and also subrecall down to the hospital/health professional level.
Quantity in commerce
9
Distribution
Product was distributed to a direct consignee/distributor in Minnesota and also directly to the only two hospital/health professional accounts involved both in Tennessee.
Product codes and lots
all codes