Z-0047-04
Brand name: Olympus ImageManager Software, Versions 6.0 through 6.4.1 (excluding version 6.3.8) Common name...
OLYMPUS Corporation of the Americas / initiated 2003-08-19 / Terminated / root cause: software
Olympus America Inc. began a recall of Brand name: Olympus ImageManager Software, Versions 6.0 through 6.4.1 (excluding version 6.3.8) Common name: Medical image storage device. Class I, 510 (k)... on . The reason FDA records is "Software malfunction; potential for simultaneous cross-linking of exam images to another patient's file". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0047-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Brand name: Olympus ImageManager Software, Versions 6.0 through 6.4.1 (excluding version 6.3.8) Common name: Medical image storage device. Class I, 510 (k) exempt. (NOTE: any ImageManager Software sold as a stand-alone system is not impacted by this software malfunction since no opportunity exists for any image cross-linking. It is a condition of the software malfunction that the observed image crosslink results from the collision of data files from two separate ImageManager computers.)
- Product code
- LMB Device, Digital Image Storage, Radiological
- Reason for recall
- Software malfunction; potential for simultaneous cross-linking of exam images to another patient's file.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- OAI sent "Urgent-Device Correction" letters, dated 8/19/03, to the end users, explaining the software malfunction. The letter also provided specific instructions to the end users until the software fix is installed. For technical assistance, users can contact OAI's Technical Assistance Center 1-800-848-9024. Accompanying the letter is a questionnaire to be completed by the end users & returned to OAI for scheduling an appointment to visit the facility to install the software fix.
- Quantity in commerce
- 1014
- Distribution
- Medical facilities (end users) nationwide and Puerto Rico, plus 5 foreign accounts located in Canada, Mexico, Ecuador and Chile.
- Product codes and lots
- Versions 6.0 through 6.4.1 (excluding version 6.3.8)
FIRST ARTICLE / WEEKLY