Z-0165-04
AEROSET Immunoglobulin M, list number 1E01-01, manufactured by Abbott Laboratories, South Pasadena, CA
Abbott Medical / initiated 2002-07-31 / Terminated
Abbott Laboratories Diagnostic Div began a recall of AEROSET Immunoglobulin M, list number 1E01-01, manufactured by Abbott Laboratories, South Pasadena, CA on . The reason FDA records is "Assay may produce falsely low IgM results without error codes when testing specimens contain elevated serum paraproteins associated with various myeloma or other malignancies". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0165-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- AEROSET Immunoglobulin M, list number 1E01-01, manufactured by Abbott Laboratories, South Pasadena, CA
- Product code
- DFT Igm, Antigen, Antiserum, Control
- Reason for recall
- Assay may produce falsely low IgM results without error codes when testing specimens contain elevated serum paraproteins associated with various myeloma or other malignancies.
- FDA root cause
- Other
- Action
- Consignees were provided with assay parameters for the RCD flag which may enhance the detection of these type of specimens. The assay parameter settings for the RCD flag were distributed to all AEROSET consignees in the United States by direct mail and to country managers worldwide via a Product Information letter dated July 31, 2002. Also provided with the Product Information letter was an AEROSET IgM Application Sheet containing the updated assay parameters for the RCD flag.
- Quantity in commerce
- 205 units
- Distribution
- Nationwide to the following states: Texas, Illinois, Indiana, Tennessee, Alaska, Pennsylvania, New York, Florida, Massachusetts, Missouri, Alabama, North Carolina, California, Ohio, Utah, Michigan, Arizona, Wisconsin, Oklahoma, Maine, Kansas, Georgia, Mississippi, Connecticut, Rhode Island, Washington, Wyoming, Louisiana, Washington, Maryland, New Jersey, Arkansas, Iowa, South Carolina, Colorado, Minnesota, North Dakota, and West Virginia. Worldwide to the following countries: Argentina, Australia, Brazil, Germany, England, Canada, Chile, Colombia, Dominican Republic, Denmark, France, Hong Kong, Japan, Korea, Singapore, Guatemala, New Zealand, Thailand, Turkey, Spain, Italy, South Africa, Sweden, Greece, Portugal, Venezuela, Pakistan, Finland, Ireland, and Taiwan.
- Product codes and lots
- All lots