Z-0173-04

AxSYM Analyzer, Small Heater Block Clamp, manufactured by Abbott Laboratories, Irving, Texas

Abbott Medical / initiated 2002-06-06 / Terminated

Abbott Laboratories, Inc began a recall of AxSYM Analyzer, Small Heater Block Clamp, manufactured by Abbott Laboratories, Irving, Texas on . The reason FDA records is "Clamp can cause misaligned seating of heater block resulting in inaccurate dispension of MEIA Wash Buffer or MUP into the reaction cell. Causes inconsistent results". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0173-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
AxSYM Analyzer, Small Heater Block Clamp, manufactured by Abbott Laboratories, Irving, Texas
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Clamp can cause misaligned seating of heater block resulting in inaccurate dispension of MEIA Wash Buffer or MUP into the reaction cell. Causes inconsistent results.
FDA root cause
Other
Action
All consignees received Technical Service Bulletin (TSB) #83-081. This TSB gave instructions on how to install the AxSYM Small Heater Block clamp upgrade implemented June 6, 2002.
Quantity in commerce
15437 units worldwide
Distribution
Nationwide and to the following foreign locations: Mexico, Viet Nam, Venezuela, Columbia, Argentina, Chile, Japan, Austria, Denmark, France, Israel, Norway, Saudi Arabia, Spain, Turkey, India, Phillippines, Brazil, Puerto Rico, Dominicam Republic, Canada, Germany, Taiwan, Thailand, Belgium, Egypt, Greece, Italy, Poland, South Africa, Sweden, United Arab Emirates, Malaysia, Singapore, Hong Kong, Australia, New Zealand, China, Uruguay, Korea, England, Czech Republic, Finland, Hungary, Netherlands, Portugal, Eastern Europe, Switzerland, Central America, and Pakistan
Product codes and lots
All Serial Numbers