Z-0172-04

ARCHITECT Processing Module, manufactured by Abbott Laboratories, Irving, Texas

Abbott Medical / initiated 2002-05-08 / Terminated

Abbott Laboratories, Inc began a recall of ARCHITECT Processing Module, manufactured by Abbott Laboratories, Irving, Texas on . The reason FDA records is "If the interlock cable connected to the door lid sensor is disconnected or fails, the lid is detected as closed even though it may be open". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0172-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
ARCHITECT Processing Module, manufactured by Abbott Laboratories, Irving, Texas
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
If the interlock cable connected to the door lid sensor is disconnected or fails, the lid is detected as closed even though it may be open.
FDA root cause
Other
Action
Firm sent technical service bulletins with corrective instructions to consignees with all existing ARCHITECT i 2000 processing modules in February 2003.
Quantity in commerce
1,167 units worldwide
Distribution
States of Virginia, Illinois, New York, Texas, Oklahoma, Pennsylvania, ILlinois, Michigan, California, New Hampshire, Ohio, Florida, Maryland, Nevada, Georgia, Minnesota, Colorado, Arizona, Missouri, Louisiana, Oregon, Wisconsin, Hawaii, Connecticut, New Hampshire, Massachusetts, Washington, South Carolina, Mississippi, Alabama, and Tennessee, and the following foreign countries: Austria, Denmark, Germany, Italy, Poland, South Africa, Sweden, England, Colombia, Australia, Japan, New Zealand, Belgium, Finland, Greece, Netherlands, Portugal, Romania, Turkey, Canada, Mexico, Hong Kong, Korea, Taiwan, Czech Republic, France, Israel, Norway, Saudi Arabia, Spain, United Arab Emirates, Brazil, Venezuela, India, Malaysia, and Thailand.
Product codes and lots
All Serial numbers