Z-0038-04

Sideport Kit w/Oxygen Line, consisting of an adjustable sideport valve, an O2 port elbow and O2 supply...

Halyard Health / initiated 2003-09-10 / Terminated

VIASYS Med Systems began a recall of Sideport Kit w/Oxygen Line, consisting of an adjustable sideport valve, an O2 port elbow and O2 supply tubing; Mfg. for Corpak MedSystems, Wheeling, IL 60090... on . The reason FDA records is "The exhalation hole in the sideport relief valve may be blocked, allowing excess pressure build-up". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0038-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Halyard Health
Product
Sideport Kit w/Oxygen Line, consisting of an adjustable sideport valve, an O2 port elbow and O2 supply tubing; Mfg. for Corpak MedSystems, Wheeling, IL 60090; 20 units per case; reorder #BLD-83167
Product code
CAI Circuit, Breathing (W Connector, Adaptor, Y Piece)
Reason for recall
The exhalation hole in the sideport relief valve may be blocked, allowing excess pressure build-up.
FDA root cause
Other
Action
VIASYS telephoned the direct accounts and sent out follow-up faxes on 9/10-11/03, informing them of the potential for a blocked exhalation hole in the sideport relief valve cap which may result in excess pressure build-up. The accounts were instructed to check their inventory for the affected catalog numbers and lots numbers, place any found on hold for return to VIASYS, and sub-recall the affected product from their customers.
Quantity in commerce
1840 units
Distribution
Missouri, Texas, New York, Illinois, Kansas, California, Pennsylvania, Georgia, Arizona, Massachusetts, Maine, Oklahoma and Virginia.
Product codes and lots
reorder #BLD-83167, lots 11385 and 11541