Z-0039-04
Coulter AcT 5diff Open Vial Hematology Analyzers Part Numbers: 6605580, 6605581, 6605604
Siemens Healthcare Diagnostics Inc. / initiated 2003-09-08 / Terminated
Beckman Coulter Inc began a recall of Coulter AcT 5diff Open Vial Hematology Analyzers Part Numbers: 6605580, 6605581, 6605604 on . The reason FDA records is "Samples with high fat content will not count platelets accurately". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0039-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Coulter AcT 5diff Open Vial Hematology Analyzers Part Numbers: 6605580, 6605581, 6605604
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- Samples with high fat content will not count platelets accurately.
- FDA root cause
- Other
- Action
- Firm issued an advisory letter to users to retain as part of the Quality Manual for the instruments. It cautions to check specimens for interfering lipemia. Letter was sent 9/8/2003.
- Quantity in commerce
- 36
- Distribution
- United States and Canada
- Product codes and lots
- All instruements are affected.