Z-0040-04

Coulter AcT 5diff Cap Pierce Hematology Analyzers Part Numbers: 6605641, 6605705

Siemens Healthcare Diagnostics Inc. / initiated 2003-09-08 / Terminated

Beckman Coulter Inc began a recall of Coulter AcT 5diff Cap Pierce Hematology Analyzers Part Numbers: 6605641, 6605705 on . The reason FDA records is "Samples with high fat content will not count platelets accurately". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0040-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Coulter AcT 5diff Cap Pierce Hematology Analyzers Part Numbers: 6605641, 6605705
Product code
GKZ Counter, Differential Cell
Reason for recall
Samples with high fat content will not count platelets accurately.
FDA root cause
Other
Action
Firm issued an advisory letter to users to retain as part of the Quality Manual for the instruments. It cautions to check specimens for interfering lipemia. Letter was sent 9/8/2003.
Quantity in commerce
451
Distribution
United States and Canada
Product codes and lots
None apply.