Western Medica OPC Portable Oxygen Conserving Regulator, Model LIN-7005, 1/2 - 5 LPM, packaged under the...

Western Enterprises, Inc. / initiated 2003-10-02 / Terminated

Western a Scott Fetzer Co began a recall of Western Medica OPC Portable Oxygen Conserving Regulator, Model LIN-7005, 1/2 - 5 LPM, packaged under the Lincare label on . The reason FDA records is "The demand valve is defective and does not function as intended in the 'Conservative Mode'". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0055-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Western Enterprises, Inc.
Product
Western Medica OPC Portable Oxygen Conserving Regulator, Model LIN-7005, 1/2 - 5 LPM, packaged under the Lincare label.
Product code
BZD Ventilator, Non-Continuous (Respirator)
Reason for recall
The demand valve is defective and does not function as intended in the 'Conservative Mode'.
FDA root cause
Other
Action
The firm notified the affected consignees by letter on 10/2/2003. Customers were asked to return the units for replacements.
Quantity in commerce
1,537 units
Distribution
The device was distributed nationwide and into China and Japan.
Product codes and lots
Unit S/Ns beginning with ''DHE0010xxxxx'' and ''DHE001100001'' thru ''DHE001102881''