Z-0557-04

The device is SCIPRO Flexible Stone Dislodger and Retrieval Set, Catalog # SP3000146-ST used for removal of...

Scion Cardio-Vascular, Inc. / initiated 2003-06-30 / Terminated

Scion Cardio-Vascular, Inc. began a recall of The device is SCIPRO Flexible Stone Dislodger and Retrieval Set, Catalog # SP3000146-ST used for removal of calculi, fragments of medical devices and other... on . The reason FDA records is "The firm received field complaints that it was difficult to open the stone retrieval basket during use". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0557-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Scion Cardio-Vascular, Inc.
Product
The device is SCIPRO Flexible Stone Dislodger and Retrieval Set, Catalog # SP3000146-ST used for removal of calculi, fragments of medical devices and other foreign objects uring urological or gatroenterological procedures. The device has the manufacturer listed on the label which is Scion Cardio-Vascular, Inc., 14256 SW 119 Avenue, Miami, FL 33186.
Product code
FGO Dislodger, Stone, Flexible
Reason for recall
The firm received field complaints that it was difficult to open the stone retrieval basket during use.
FDA root cause
Other
Action
The firm mailed a recall letter to there consignees on June 30, 2003 requesting return of any unused devices. The firm has received 100% response and accounted for all recalled devices.
Quantity in commerce
41
Distribution
The software systems were installed at 2 medical centers in Illinois and Alabama.
Product codes and lots
Lot Number OML30522-01