Z-0081-04

Sonotrode Probe, a straight nephroscopy probe with a 3.5 mm diameter and 362 mm working length; an accessory...

Richard Wolf Medical Instruments Corp. / initiated 2003-10-09 / Terminated

Richard Wolf Medical Instrument Corp began a recall of Sonotrode Probe, a straight nephroscopy probe with a 3.5 mm diameter and 362 mm working length; an accessory to the Richard Wolf Model 2271 Ultrasound... on . The reason FDA records is "The sonotrode tube could become blocked by the stone fragments, causing overheating and damage to the device". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0081-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Richard Wolf Medical Instruments Corp.
Product
Sonotrode Probe, a straight nephroscopy probe with a 3.5 mm diameter and 362 mm working length; an accessory to the Richard Wolf Model 2271 Ultrasound Generator Unit 2271; Richard Wolf Medical Instruments Corp., Vernon Hills, IL 60061
Product code
FFK--
Reason for recall
The sonotrode tube could become blocked by the stone fragments, causing overheating and damage to the device.
FDA root cause
Other
Action
Recalled from the hospitals by sales representative telephone calls on 10/9/03 and follow-up visits to retriev the recalled lot and replace the units with a new lot of sonotrodes.
Quantity in commerce
10 probes
Distribution
Nebraska, Oregon, Minnesota and Illinois.
Product codes and lots
Model 8963.535, lot number M000580