Z-0288-04

Radiant Heater Door Vent for Ohmeda Medical''s Giraffe (tm) OmniBed neonate bed

GE Healthcare / initiated 2003-10-17 / Terminated

Ohmeda Medical, A Div of Datex-Ohmeda , Inc began a recall of Radiant Heater Door Vent for Ohmeda Medical''s Giraffe (tm) OmniBed neonate bed on . The reason FDA records is "Heater vent door on neonatal incubator may malfunction". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0288-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Radiant Heater Door Vent for Ohmeda Medical''s Giraffe (tm) OmniBed neonate bed.
Product code
FMZ Incubator, Neonatal
Reason for recall
Heater vent door on neonatal incubator may malfunction.
FDA root cause
Other
Action
The recalling firm notified domestic consignees by certified letter and international distributors by e-mail dated 10/17/2003. The recall notice was incorporated into a firm''s follow-up letter concerning an on-going product correction of the elevating base on the same neonatal beds. The letter advises of the malfunctioning heater door vent and states that the firm will replace all existing vent doors with doors made of higher temperature rated material. Firm will provide retrofit kits for consignee or service representative installation.
Quantity in commerce
1404 beds
Distribution
Products were sold to 365 direct accounts around the US and 58 international distributors.
Product codes and lots
HDGDyyyyy HDGEyyyyy HDGFyyyyy HDGGyyyyy where ''yyyyy'' represents any number less than 60000