RITA brand ThermoPad, Electrosurgical RF Generator and accessories Part Number: 700-102649

Boston Scientific Corporation / initiated 2003-10-06 / Terminated

Rita Medical Systems began a recall of RITA brand ThermoPad, Electrosurgical RF Generator and accessories Part Number: 700-102649 on . The reason FDA records is "The product displays inaccurate temperature readings, patient skin burns can occur before getting to higher temperature reading". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0087-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
RITA brand ThermoPad, Electrosurgical RF Generator and accessories Part Number: 700-102649
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
The product displays inaccurate temperature readings, patient skin burns can occur before getting to higher temperature reading.
FDA root cause
Other
Action
On 10/6/03, the firm issued letters via Federal Express to all its direct consignees, informing them of the affected devices and providing instructions on the recall.
Quantity in commerce
137 + 225 units (lot Y082003)
Distribution
Product was released for distribution to 66 consignees in US. The product received nationwide distribution. The recall was appropriately extended to the user level; i.e., the hospitals that received the recalled product. There were no shipments to foreign or Canadian distribution, however, it includes one military distribution: Wilford Hall Medical Center, 2200 Berquist Dr., 31, Lackland AFB, TX 78236.
Product codes and lots
Lot numbers: Y072103-3, Y082003