Z-0244-04

NEONATE 4 DISPOSABLE BLOOD PRESSURE CUFF WITH 1 TUBE, a disposable one piece cuff, for single patient use...

Welch Allyn Inc / initiated 2003-10-27 / Terminated

Welch Allyn Inc began a recall of NEONATE 4 DISPOSABLE BLOOD PRESSURE CUFF WITH 1 TUBE, a disposable one piece cuff, for single patient use only. Sold under the Welch Allyn ''Tycos'' and... on . The reason FDA records is "Incorrect location of inflatable portion of cuff that may result in cuff falling off patient's arm, error code when used with patient monitor, or an inaccurate reading". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0244-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Welch Allyn Inc
Product
NEONATE 4 DISPOSABLE BLOOD PRESSURE CUFF WITH 1 TUBE, a disposable one piece cuff, for single patient use only. Sold under the Welch Allyn ''Tycos'' and Allegiance ''Tactics'' labels. The ''4'' represents the cuff size. (4 = 6.9 to 11.7 cm). These are sold in various configurations and assigned different part numbers. Devices are labeled as follows: Welch Allyn Tycos #5082-104-1 (10 pack/1 tube); Welch Allyn Tycos #5082-241-9 (5 pack/1 tube, multi-size pack); Allegiance Tactics #30502-113S (case of 40/1tube); Welch Allyn Tycos/Allegiance #5082-241-15 (5 pack/1 tube, multi-size pack); Firm on label: (1) Welch Allyn Tycos and Tycos/Allegiance label (package insert and shipping case only): ''Welch Allyn, Inc., Skaneateles Falls, NY''; (2) Allegiance Tactics label ''Distributed by: Allegiance Healthcare Corporation, McGaw Park, IL 60085-6787 USA''
Product code
DXQ Blood Pressure Cuff
Reason for recall
Incorrect location of inflatable portion of cuff that may result in cuff falling off patient's arm, error code when used with patient monitor, or an inaccurate reading.
FDA root cause
Other
Action
Letters dated October 27, 2003 to wholesalers/ distributors with instructions to remove/return product and notify end users.
Quantity in commerce
26190
Distribution
The product was sold thru 11 distributors. The product was shipped to end users nationwide. There is one foreign distributor in Singapore. In addition, Welch Allyn shipped product to their affiliates in Ireland, Australia, and South Africa.
Product codes and lots
Code dates between June 20, 2003 and October 8, 2003. (Codes found on the outer carton only.)