IntelliSystem25 Inflation Device and Fluid Dispensing Syringe - Label on package reads in part...
Merit Medical Systems, Inc began a recall of IntelliSystem25 Inflation Device and Fluid Dispensing Syringe - Label on package reads in part: ''INTELLISYSTEM 25 WITH MAP152 MAP152 CONTAINS: MERIT ACCESS... on . The reason FDA records is "There is a potential for the PETG blister trays to bond together with varying degrees of attachment. The potential for non-sterility exists when the attached trays are seperated by force and holes or cracks are...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0138-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- IntelliSystem25 Inflation Device and Fluid Dispensing Syringe - Label on package reads in part: ''INTELLISYSTEM 25 WITH MAP152 MAP152 CONTAINS: MERIT ACCESS PLUS LARGE BORE HEMOSTASIS VALVE, TORGUE DEVICE AND METAL GUIDE WIRE INSERTION TOOL...MERIT MEDICAL'', Latex free. Sterile if package is unopened or undamaged. Part # 1525/B
- Product code
- DXT Injector And Syringe, Angiographic
- Reason for recall
- There is a potential for the PETG blister trays to bond together with varying degrees of attachment. The potential for non-sterility exists when the attached trays are seperated by force and holes or cracks are generated in the seperated trays.
- Root cause tag
- packaging
- FDA root cause
- Other
- Action
- Merit notified sales representatives, domestic and international accounts and dealers immediately by letter dated October 3, 2003. The manufacturer requests that the use of the affected inventory should immediately cease and that on-hand units should be promptly isolated from general inventory for simultaneous inspection by a site representative and a Merit sales representative.
- Quantity in commerce
- 626 units
- Distribution
- Distributed nationwide. International distribution includes Canada, Japan, Korea, England and France.
- Product codes and lots
- Lot # A285003.