Z-0237-04

3M Attest 1264P Biological Indicators, in boxes of 25 per box

3M Company / initiated 2003-10-24 / Terminated / root cause: sterility

3m Health Care began a recall of 3M Attest 1264P Biological Indicators, in boxes of 25 per box on . The reason FDA records is "The 3M Attest 1264/1264P Biological Indicators of ETO sterilization contain a microbiological contaminant which can affect the performance of the positive control and the indicators to some limited extent. The color on...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.

Recall record

Recall number
Z-0237-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
3M Company
Product
3M Attest 1264P Biological Indicators, in boxes of 25 per box
Product code
FRC Indicator, Biological Sterilization Process
Reason for recall
The 3M Attest 1264/1264P Biological Indicators of ETO sterilization contain a microbiological contaminant which can affect the performance of the positive control and the indicators to some limited extent. The color on a positive control may revert to negative after 24 hours of incubation.
Root cause tag
sterility
FDA root cause
Other
Action
The firm conducted a field correction which began on 10/24/03. They provided revised labeling for the product. Letters, dated October 24, 2003, were sent to consignees. The letters had a description of the problem, instructions for doing the relabeling in the field, and recommendations for users of the product.
Quantity in commerce
1844 boxes of 25 biological indicators per box
Distribution
The product was distributed nationwide in the United States and worldwide.
Product codes and lots
Lot numbers 2005-07 DW, 2005-08 DD, and 2005-08 DL