Z-0237-04
3M Attest 1264P Biological Indicators, in boxes of 25 per box
3M Company / initiated 2003-10-24 / Terminated / root cause: sterility
3m Health Care began a recall of 3M Attest 1264P Biological Indicators, in boxes of 25 per box on . The reason FDA records is "The 3M Attest 1264/1264P Biological Indicators of ETO sterilization contain a microbiological contaminant which can affect the performance of the positive control and the indicators to some limited extent. The color on...". As of , the recall status on record is Terminated, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-0237-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- 3M Company
- Product
- 3M Attest 1264P Biological Indicators, in boxes of 25 per box
- Product code
- FRC Indicator, Biological Sterilization Process
- Reason for recall
- The 3M Attest 1264/1264P Biological Indicators of ETO sterilization contain a microbiological contaminant which can affect the performance of the positive control and the indicators to some limited extent. The color on a positive control may revert to negative after 24 hours of incubation.
- Root cause tag
- sterility
- FDA root cause
- Other
- Action
- The firm conducted a field correction which began on 10/24/03. They provided revised labeling for the product. Letters, dated October 24, 2003, were sent to consignees. The letters had a description of the problem, instructions for doing the relabeling in the field, and recommendations for users of the product.
- Quantity in commerce
- 1844 boxes of 25 biological indicators per box
- Distribution
- The product was distributed nationwide in the United States and worldwide.
- Product codes and lots
- Lot numbers 2005-07 DW, 2005-08 DD, and 2005-08 DL