Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0155-04

Midwest Tradition PB Fiber Optic High Speed Handpiece with Push Button Chuck (5 tube connection), catalog...

DentSply -Professional Division / initiated 2003-10-14 / Terminated

DentSply -Professional Division began a recall of Midwest Tradition PB Fiber Optic High Speed Handpiece with Push Button Chuck (5 tube connection), catalog #790045 and 790045T; Manufactured for DENTSPLY... on . The reason FDA records is "Small cracks in the dental handpiece's chuck could allow the dental burs to drift out of the chuck over time". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0155-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
DentSply -Professional Division
Product
Midwest Tradition PB Fiber Optic High Speed Handpiece with Push Button Chuck (5 tube connection), catalog #790045 and 790045T; Manufactured for DENTSPLY Professional, A Division of DENTSPLY International, 901 West Oakton Street, Des plaines, IL 60018-1884 USA
Product code
EFB Handpiece, Air-Powered, Dental
Reason for recall
Small cracks in the dental handpiece's chuck could allow the dental burs to drift out of the chuck over time.
FDA root cause
Other
Action
Dealers were telephoned on 10/14-16/03 and sent follow-up faxed letters on the same dates. The dealers were informed of the potential for cracked bur chucking mechanisms in the handpieces, requested to examine their stocks for the affected serial numbers, using the faxed list of serial numbers shipped to each account, and return the affected handpieces to DENTSPLY via FedEx. The dealers were also requested to verify the serial number list to those handpieces shipped to customers. If the affected serial numbers were distributed, the customer was requested to supply a list of those customers with contact information to DENTSPLY. As those lists are received by DENTSPLY, end user letters are sent to those customers, informing them of the problem with the bur chucking mechanism and requesting the retutrn of the affected serial numbers for examination and replacement of the bur chucking mechanism if necessary.
Quantity in commerce
698 units
Distribution
Nationwide, Canada and Mexico
Product codes and lots
Chuck manufacturing lot numbers 7/15/03, 7/25/03, 8/12/03, 8/14/03; handpiece serial numbers 52336288 through 52429923. Not all units in the serial number range were affected. 446 units were sold new with affected chucks and 252 units were repaired with the affected chucks.