Z-0145-04
Portable Oxygen Analyzer
Teledyne Analytical Instruments / initiated 2003-10-27 / Terminated
Teledyne Analytical Instruments began a recall of Portable Oxygen Analyzer on . The reason FDA records is "Potential failure to display the battery low light to the end-user". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0145-04
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teledyne Analytical Instruments
- Product
- Portable Oxygen Analyzer
- Product code
- CCL Analyzer, Gas, Oxygen, Gaseous-Phase
- Reason for recall
- Potential failure to display the battery low light to the end-user.
- FDA root cause
- Other
- Action
- All affected customers will be notified by letter on 10/27/2003. All units returned will be reworked and 100% retested prior to shipment.
- Quantity in commerce
- 170
- Distribution
- FL, CA, UT, AZ and MO. New Zealand, Australia, Canada and the United Kingdom
- Product codes and lots
- Model AX300, Model MX300