Portable Oxygen Analyzer

Teledyne Analytical Instruments / initiated 2003-10-27 / Terminated

Teledyne Analytical Instruments began a recall of Portable Oxygen Analyzer on . The reason FDA records is "Potential failure to display the battery low light to the end-user". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0145-04
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Teledyne Analytical Instruments
Product
Portable Oxygen Analyzer
Product code
CCL Analyzer, Gas, Oxygen, Gaseous-Phase
Reason for recall
Potential failure to display the battery low light to the end-user.
FDA root cause
Other
Action
All affected customers will be notified by letter on 10/27/2003. All units returned will be reworked and 100% retested prior to shipment.
Quantity in commerce
170
Distribution
FL, CA, UT, AZ and MO. New Zealand, Australia, Canada and the United Kingdom
Product codes and lots
Model AX300, Model MX300